InterOcean logo InterOcean systems, inc.
Home Products Technical Support Customer Service Company Profile Contact Us Company News Sitemap

Model 1431 Hydrophone

The Model 1431 Hydrophone has a broadband uniform receiving sensitivity over the range of 200 Hz to 70 kHz. It has a receiver sensitivity of -185 dB/ubar/v +/- 3 dB. The Hydrophone contains an integral mounted FET preamp and provisions are made for calibration signal insertion. The Hydrophone is neoprene encapsulated with a brass header and protective shroud.


Model 1431 Response Graph


Specifications
Dimensions: 14 cm (5.5 in) length,
2.6 cm dia. (1.0 in) phone,
6.4 cm (2.5 in) base.
Weight: 1.6 Kg (3.5 lbs) in air.
Receiving Response: -185dB re 1 V/uPa +/- 3 db from 200 Hz to 70 kHz.
Horizontal Beam Pattern: Omnidirectional +/- 1 dB.
Preamp: Integral mounted FET preamp.
Power Requirement: 12 VDC, nominal 2 ma or less current drain.
Material: Neoprene encapsulated brass header with preamp cavity.
Cable Requirement: 4 conductors standard and calibration signal conductor can be provided, if desired.


Home Products Technical Support Customer Service Company Profile Contact Us Company News Sitemap

InterOcean systems, inc.
4241 Ponderosa Avenue, Suite A, San Diego, CA 92123-6501 Phone: (858) 565-8400 Fax: (858) 268-9695
email: sales@interoceansystems.com